Leak Tester MF-200

1、Product Overview

Leak Tester MF-200 is purpose-built for vacuum leak and seal-integrity testing — as well as microbial ingress testing — of a wide range of packaging: pharmaceutical glass bottles, vials, plastic solid bottles, blister packs, aluminum-plastic sachets, syringes, pipettes, vacuum bags, cans, milk-powder bags, foil bags, eye-drop bottles, medical device packaging, liquid bags, jelly cups, caps, cosmetic bottles and daily-chemistry packaging.

2、Application Fields

Application Fields — Widely adopted by pharmaceutical manufacturers, pharmaceutical packaging producers, drug inspection centers, medical device companies, food enterprises and cosmetics companies for professional seal-integrity inspection.

3、Testing Principle

By evacuating the vacuum chamber, a pressure differential is created between the inside and outside of the sample. For a sample immersed in water, any gas escaping from it is observed; alternatively, the sample’s expansion and the recovery of its original shape after the vacuum is released are observed. Either way, the seal performance of the sample is judged accurately.

Complies with GB/T 15171, ASTM D3078 and other standards.

4、Key Features

★Precision push-button control — operation made effortless.

★High-clarity acrylic vacuum chamber — clear, tough and durable.

★Automatic constant-pressure air replenishment keeps the whole test at the preset vacuum level.

★Quality vacuum components deliver stable, long-lasting performance.

★Adjustable holding (dwell) time.

5、Technical Specifications

ParameterSpecification
Vacuum Setting Range0 ~ -90 KPa
Effective Size of Vacuum ChamberΦ270 mm × 200 mm
Measuring Time≤ 9999 s
Time Accuracy1 s
Pumping Rate2 KPa·L/s
Vacuum Pump Power150 W
Main Unit Dimensions300 mm × 240 mm × 125 mm
Operating VoltageAC 220V, 50Hz
Total Weight 20 kg